ISO/IEC 17025 Calibration
- Per B89.4.1b-2001 and OEM specifications.
- Measurement uncertainty for all tests.
- Before and after results.
- Repeatability per B89.
- X, Y, and Z linear accuracy per B89.
- Squareness / Perpendicularity.
- Volumetric accuracy per B89 if OEM specifies.
- Linear temperature compensation to 68F°.
- Approved test sheets.
- Approved preliminary and final certificates.
- Approved calibration sticker.
- Hard measurement data.
- Meets A2LA traceability policy.
- The quality manager reviews the final certificate.
- Meets QS9000, ISO, and ISO/TS requirements.
- Adjustments are made to meet OEM specifications.
Includes preventive maintenance after the before measurements.
- Clean and inspect fans, filters, airlines, scales, and way surfaces.
- Check and adjust read heads and signals.
- Inspect and adjust fine feeds, air / hard bearings, counterbalance, machine level, and failsafe.
- Inspect power supplies for AC ripple and proper voltage.
- Recommendations will be made for repairs if needed.
-
EOC Calibration (Not Accredited)
- Before and after results.
- Repeatability per B89 X, Y, and Z linear accuracy per B89.
- Squareness / Perpendicularity
- Volumetric accuracy (15 position ball-bar) if OEM specifies.
- Hard measurement data.
- Calibration sticker
- Calibration certificate
- Traceable to N.I.S.T
- Adjustments are made to meet OEM Spec.
Includes preventive maintenance after the before measurements.
- Clean and inspect fans, filters, airlines, scales, and way surfaces.
- Check and adjust read heads and signals.
- Inspect and adjust fine feeds, air / hard bearings, counterbalance, machine level, and failsafe.
- Inspect power supplies for AC
- Ripple and proper voltage.
- Recommendations will be made for repairs if needed.
Servo Option
Full Servo preventive maintenance includes:
- Inspect motors.
- Inspect belts and alignment of drives.
- Check bearing wear
- Check KV parameters
- Check positioning
- Check balance
- Check slip
- Check overshoot
- Belt tension
- Adjustments will be made if needed to the inspected items.
- Recommendations will be made for repairs if needed.
-
Factory Calibration (Not Accredited)
- No before measurements.
- Repeatability per B89 X, Y, and Z linear accuracy per B89.
- Squareness / Perpendicularity
- Volumetric accuracy (15 position ball-bar) if OEM specifies.
- Hard measurement data.
- Calibration sticker
- Calibration certificate
- Traceable to N.I.S.T
- Adjustments are made to meet OEM specifications.
Includes preventive maintenance before measurements are taken.
- Clean and inspect fans, filters, airlines, scales, and way surfaces.
- Check and adjust read heads and signals.
- Inspect and adjust fine feeds, air / hard bearings, counterbalance, machine level, and failsafe.
- Inspect power supplies for AC ripple and proper voltage.
- Recommendations will be made for repairs if needed.
Z Roll Option
- Z-Axis Roll Performance Test.
- Recommended for customers who utilize long offset probe extensions.